| Patient Praperation | ||
IMAGING PARAMETERS |
||
| Plane | Sagittal |
Coronal |
Mode |
2D |
3D |
Pulse Seq |
Spin Echo |
Vasc TOF SPGR |
| Imaging Options | RC,NPW |
Fast, MPh, ZIP2 |
SCAN TIMING |
||
| # of Echoes | 1 |
1 |
| TE | Min Full |
Minimum |
| TR | 385 |
-- |
| Flip Angle | -- | 60 (45-90) |
| Bandwidth | 16 |
31.25 |
| SAT | None |
None |
ACQUISITION TIMING |
||
| Freq | 256 |
256 |
| Phase | 128 |
192 (128-256) |
| NEX | 2 |
0.50(5-1) |
| Phase FOV | -- | 1 |
| Locs Before Pause | -- | 1 |
| Freq DIR | S/I |
S/I |
| Auto Center Freq | Peak |
Water |
| Auto Shim | Yes |
Yes |
| Contrast | No |
Dilute Gd |
SCANNING RANGE |
||
| FOV | 44(36-48) |
40(32-48) |
| Slice Thickness | 10 |
3(2-5) |
| Spacing | Interleave |
-- |
| Start - End | L145-R145 |
|
| # Slices | 30 |
40(30-50) |
| Scan Time (min:sec) | -- | -- |
Indications:
Patient Preparation: start intravenous line in the arm on the same side as the subclavian vein that is to be imaged. Connect SmartSet intrvenous tubing (www.topspins.com, 734-623-6400). If both left and right subclavian veins are to be imaged or if imaging the SVC is essential, than start bilateral intravenous lines. The intravenous site should be located preferably in the antecubital fossa. Intravenous sites in the forearm, wrist or hand are also acceptable provided they can tolerate a flow rate of 2ml/sec. Any size angiocatheter down to 22 gauge is acceptable. Smaller angiocatheters (larger gauge number) may have too much resistance to allow a sufficient rate of contrast injection.
After starting the intravenous lines, place cushions under both forearms and elbows to raise up the arms so they will be included in the relatively anterior imaging volume required to encompass the subclavian veins.
Dilution of Gadolinium: Mix one bottle of gadolinium (10 or 20 ml) into a 250ml bag of normal saline. It is acceptable to use sterile water or D5W or D5 1/2NS in patients who are salt restricted. This is especially important in patients with a history of congestive heart failure. Fill four, 60ml syringes with the diluted gadolinium contrast and attach two loaded syringes to each SmartSet.
Patient Position: Feet first, supine,
Landmark: mid sternum or mid-nipple line. Landmark lower if you need to cover more of the arm veins.

Coil: If imaging just one arm, use the torso array coil positioned over the shoulder on the side of interest and extend medial enough to cover the SVC.If bilateral, use the body coil.
SERIES 1: Sagittal T1 spin echo localizer.
Adjust the field-of-view so as the cover at least up to the level of the thyroid. The necessary size field-of-view can be determined by measuring from the landmark site up to the thyroid cartilage. In thin patients some time can be saved by using ¾ field of view.
SERIES 2: Coronal 3D MR Venography with injection of dilute gadolinium

This sequence produces a venogram image which is comparable to venography performed using X-rays and iodinated contrast. It has the advantage of using gadolinium which is completely safe and can be diluted 10 to 20-fold so that a large volume of dilute Gd will distend the venous system and demonstrate venous anatomy over a large field of view.
Filming: Use windows workstation to obtain subvolume MIPs of upper extremity veins in AP and oblique projections.
Billing: MRA Chest 771555
Computer Recontructions 76375
ICD9 codes
Sample Normal Dictation:
The patient was imaged with a Sagittal T1 weighted sequence and a Coronal 3D spoiled gradient echo pulse sequence during infusion of dilute gadolinium. Following acquisition of image data further analysis was performed on a computer workstation to obtain reformations and MIPs of the major vessels.
The axillary, subclavian and brachiocephalic veins and SVC are widely patent. No significant narrowing or thrombus is identified.
Impression:
Normal SVC and axillary, subclavian and brachiocephalic veins bilaterally.